QA-QC Protocols and Spectra

Introduction

This page contains spectra that are representative of the normal level of performance of this instrument.  A brief description of the acquisition protocol is given for each set of measurements.  More detailed information about the proper instrument set-up and operational parameters necessary to reproduce these measurements can be found in the appropriate training manual or application note.

Tuning and Calibration - EI mode - METRI

Allow the instrument to warmup for at least fifteen minutes prior to tuning and calibration.  Introduce approximately 1microliter of METRI into the reference inlet.  On the GCT page, zoom in on m/z 285.  Tune the instrument.  Typical results are shown below:

resolution

Acquire and calibrate.  Typical results are:

Typical Results

After calibration, and with m/z 285 lock mass enabled, typical METRI single scan data:

after calibration

After calibration, and with m/z 285 lock mass enabled, typical METRI 10-scan averaged data:

averaged results

Note that the single scan data is almost as accurate as the co-added data.  This is generally true when the ion statistics are favorable.  Where ion statistics are poor, co-added data will give better results.

Typical measurement errors, 10 scans co-added
Mass (TheoreticalMass (Measured)Error (PPM)
68.995268.99494
75.999875.9999-1
102.0030102.0031-1
121.0014121.0016-2
189.9966189.9969-2
265.9964265.9984-8

 

Calibration - EI mode - HEPTACOS

Tune with METRI as discussed above.  Remove the METRI.  Introduce one microliter of heptacosamine (HEPTA). Acquire data and calibrate.

calibrate
 

Typical single-scan data of HEPTA:

HEPTA scan

Sensitivity - EI mode - benzophenone

The benzophenone test mixture concentration should be approximately 5.0 nanograms per microliter (5.0 ppm).  One microliter is injected via the autosampler at a 50:1 split, resulting in approximately 100 picograms on the column.  Data is acquired isothermal at 200 degrees C; 2 scans per second.  Typical GC and MS results are shown below:

Benzo

benzo spectrum

GC Testmix - EI mode

The standard GC test mixture has 8 constituents at approximate concentrations of 50 ppm per component.  One microliter is injected via the autosampler at a 50:1 split, resulting in approximately 1 nanogram per component on the column.  Data is acquired isothermal at 110 degrees C; 2 scans per second.  Typical GC and MS results are shown below:

test mix

test mix 1

test mix 2

test mix 3

test mix 4

test mix 5

test mix 6

test mix 7

Tuning and calibration - CI positive ion

Sensitivity - CI positive ion

Tuning and calibration - CI negative ion

Sensitivity - CI negative ion