QA-QC Protocols

Introduction

The instrument may be operated in either CHNS or O mode. For CHNS, combustion and reduction happen in one tube.  

  • Tubes are packed in-house using reagents purchased from CE Elantech or EA Consumables
  • Quartz tubes are reused when possible; new tubes are purchased from CE Elantech
  • Pressed tin capsules (9mm x 5mm) are used for most samples.
  • Silver capsules (9mm x 5mm) are used for halogenated samples and samples prepared for O analysis. 
  • For volatile, liquid, or air-sensitive samples, we have a capsule sealer, which uses smooth-walled tin capsules
  • BBOT is typically used as the calibration standard. 
  • Vanadium pentoxide is used as a combustion aid, especially for samples that require accurate sulfur determination.
  • UHP grade oxygen is used to support combustion.
  • UHP grade helium is used as the carrier gas.

Instrument Preparation

The system is checked for leaks whenever a tube is replaced, or just before starting a run.

CHNS Operation

Normal operating procedure includes running a bypass, two blanks, three standards, and two "checks" (standards run as unknowns) before running submitted samples. 2,5-Bis(5-tert-butyl-benzoxazol-2-yl)thiophene (BBOT) is used as the standard for most studies. Acetanilide (ace) is another good standard, but does not contain S. The normal run sequence is:

Run #CompoundRun as...Purpose
1BBOT (unweighed)bypasscondition the system
2empty tin capsulebypassensure no BBOT remains in system 
3empty tin capsuleblankset initial blank values
4-6BBOT (2 - 3 mg)standardgenerates calibration curve
7-8BBOT (2 - 3 mg)unknown"checks" to ensure good mass %
9empty tin capsulebypassensure no BBOT remains in system
10-?user samplesunknownunknown samples
?BBOT (2 - 3 mg)unknownoptional: addt'l "checks" during run
200change tube combustion tube good for 200 samp

 

CHNS/O Reactor Packing and Configuration Without Crucible

Parameters

CHNS (Left Furnace)/Oxygen (Right Furnace)

CHNS  (Oven at 100°C for Oxygen)

Flash 200/FlashSmart Method

Flash 2000/FlashSmart Method

Data Processing

 After a run is complete, a calibration curve and chromatograms can be viewed for data processing. The calibration curve is based on the three standards and is what all unknown sample results are based off of. This should be checked whenever a new table is started to ensure accurate results. The following example calibration curve should be expected:

Nitrogen calibration curve

The results are summarized into a table to make comparison easy between each sample run. When comparing the results for a sample, it is important to look at the % relative standard deviation. This should remain below 3% for results to be reliable, and if it is not, or the samples seem off, then the chromatograms and component tables can be adjusted to try to resolve this issue. Chromatograms are used to view the peak window for each element in which the system integrates to determine the results. An example chromatogram and its component table looks like the following:

chart

The verticle red lines represent the peak window and can be adjusted in the component table to adjust integration. 

It is critical that duplicates be run in order to identify those instances of poor precision; whenever possible a third replicate is then run and the outlier (always lower than the others) is excluded from the average.